Input Specifications

All submitted imaging data must comply with the input specifications detailed below. Any datasets that fail to meet these requirements will be deemed unsuitable for processing and rejected.

The required MRI sequence is a diffusion-weighted, single-shot spin echo-planar imaging (EPI) acquisition with whole-brain coverage.

MRI Acquisition Parameters

ParameterValue
Field strength, T3
Repetition time (TR), ms6000 – 17000
Echo time (TE), ms55 – 100
In-plane resolution, mm²1.85 – 2.15
Slice thickness, mm2
Slices, n64 – 86
Slice acquisitionInterleaved
Slice gap, mm0
Flip angle, degrees90
b-values, s/mm²0, 1000
b0 images, n≥ 2
Directions, n≥ 30

Supported DICOM Transfer Syntaxes

UIDName
1.2.840.10008.1.2Implicit VR Little Endian
1.2.840.10008.1.2.1Explicit VR Little Endian
1.2.840.10008.1.2.2Explicit VR Big Endian (Retired)
1.2.840.10008.1.2.1.99Deflated Explicit VR Little Endian
1.2.840.10008.1.2.4.70JPEG Lossless, Non-Hierarchical, 1st-Order Prediction
1.2.840.10008.1.2.4.80JPEG-LS Lossless
1.2.840.10008.1.2.4.90JPEG 2000 Lossless Only
1.2.840.10008.1.2.4.91JPEG 2000 (Lossless Only)
1.2.840.10008.1.2.5RLE Lossless

Supported DICOM Tags

DICOM TagValue
Patient Age (0010, 0010)Age of patient (e.g., "060Y", "048Y")
Patient Sex (0010, 0040)Sex of patient (e.g., "F", "M")
Sender AE Title (0008, 0055)Origin identifier (Application Entity title)
neuropacs™ Products (0019, 0021)neuropacs™ product list