Input Specifications
All submitted imaging data must comply with the input specifications detailed below. Any datasets that fail to meet these requirements will be deemed unsuitable for processing and rejected.
The required MRI sequence is a diffusion-weighted, single-shot spin echo-planar imaging (EPI) acquisition with whole-brain coverage.
MRI Acquisition Parameters
Parameter | Value |
---|---|
Field strength, T | 3 |
Repetition time (TR), ms | 6000 – 17000 |
Echo time (TE), ms | 55 – 100 |
In-plane resolution, mm² | 1.85 – 2.15 |
Slice thickness, mm | 2 |
Slices, n | 64 – 86 |
Slice acquisition | Interleaved |
Slice gap, mm | 0 |
Flip angle, degrees | 90 |
b-values, s/mm² | 0, 1000 |
b0 images, n | ≥ 2 |
Directions, n | ≥ 30 |
Supported DICOM Transfer Syntaxes
UID | Name |
---|---|
1.2.840.10008.1.2 | Implicit VR Little Endian |
1.2.840.10008.1.2.1 | Explicit VR Little Endian |
1.2.840.10008.1.2.2 | Explicit VR Big Endian (Retired) |
1.2.840.10008.1.2.1.99 | Deflated Explicit VR Little Endian |
1.2.840.10008.1.2.4.70 | JPEG Lossless, Non-Hierarchical, 1st-Order Prediction |
1.2.840.10008.1.2.4.80 | JPEG-LS Lossless |
1.2.840.10008.1.2.4.90 | JPEG 2000 Lossless Only |
1.2.840.10008.1.2.4.91 | JPEG 2000 (Lossless Only) |
1.2.840.10008.1.2.5 | RLE Lossless |
Supported DICOM Tags
DICOM Tag | Value |
---|---|
Patient Age (0010, 0010) | Age of patient (e.g., "060Y", "048Y") |
Patient Sex (0010, 0040) | Sex of patient (e.g., "F", "M") |
Sender AE Title (0008, 0055) | Origin identifier (Application Entity title) |
neuropacs™ Products (0019, 0021) | neuropacs™ product list |