Input Specifications

All submitted imaging data must comply with the input specifications detailed below. Any datasets that fail to meet these requirements will be deemed unsuitable for processing and rejected.

The required MRI sequence is a diffusion-weighted, single-shot spin echo-planar imaging (EPI) acquisition with whole-brain coverage.

ParameterValue
Field strength, T3
Repetition time (TR), ms6000 – 17000
Echo time (TE), ms55 – 100
In-plane resolution, mm²1.85 – 2.15
Slice thickness, mm2
Slices, n64 – 86
Slice acquisitionInterleaved
Slice gap, mm0
Flip angle, degrees90
b-values, s/mm²0, 1000
b0 images, n≥ 2
Directions, n≥ 30

Last updated: June 24, 2025